Spectrum Analyzer System |
|
|
- Reliable laser modal analysis from 275 nm to 1625 nm
- Confocal configuration for easy alignment
- Choice of 0.3,1.5,7.5 or 30 GHz FSR (Free Spectral Range)
- User-interchangeable mirrors
- New updated controller
|
| Specifications |
1.5 GHz Free Spectral Range |
7.5 GHz Free Spectral Range |
30 GHz Free Spectral Range |
300 MHz Free Spectral Range |
| Cavity Type: |
5 cm confocal
|
1 cm confocal
|
1/4 cm confocal
|
25 cm confocal
|
| Finesse: |
>200
|
>200
|
>100
|
>200
|
| Instrumental Bandwidth: |
7.5 MHz
|
37.5 MHz
|
300 MHz
|
1.5 MHz
|
| Aperature: |
7.5
|
4.3 mm
|
2.0 mm
|
10.0 mm
|
 |
| Piezoelectric Motion: |
~4 nm/volt
|
| Volts per FSR at 600 nm (nominal): |
37.5 volts
|
| Maximum Voltage: |
500 VDC
|
| Crystal Capacitance: |
0.04 m F |
| Maximum Sweep Rate: |
1-2 msec
|
| Scan Non-linearity: |
2% at 100 VDC
|

|
WaveMaster Laser Wavelength Meter
Pulse and CW Wavelength Measurement |
 |
|
- 0.005 nm accuracy
- 0.001 nm resolution
- Internal self-calibration
- Easy setup
- No warm up required
- 380-1095 nm wavelength range
- RS-232 and Optional GPIB Interfaces
|
| Specifications |
| Wavelength Coverage: |
380-1095 nm
|
| Accuracy: |
0.005 nm
|
| Resolution: |
0.001 nm
|
| Min. Pulse Rep Rate: |
single shot
|
| Max. Pulse Rep Rate: |
CW
|
| Max. Signal Bandwidth: |
2 nm @ 400 nm
|
| 3 nm @ 600 nm
|
| 5 nm @ 1000 nm
|
| Min. Signal: |
20 µW CW @ 632 nm |
| 2 mJ pulsed @ 1064 nm
|
| Max. Signal: |
100 mW CW @ 632 nm |
| 100 mJ pulsed @ 1064 nm
|
| Display Update: |
3 Hz
|
| Size: |
281 mm wide x 105 mm high x 352 mm deep
|
| Storage Conditions: |
-10 °C to 50 °C
|
| Relative humidity: |
non-condensing and <80% |
| Shock: |
<4 G |
| Use Conditions: |
-10-40 °C
|
| Relative humidity: |
non-condensing and <80 |
| Shock: |
<4 G |
| Power Supply (supplied): |
Universal 90-250 VAC, 40-72 Hz
|
|
WaveMate
Wavelength Meter |
|
|
- Fast, cost-effective, wavelength measurements
- CW and pulsed lasers
- High sensitivity: 20 µW CW, 10 µJ pulsed
- Excellent for diode lasers
|
| Specifications |
| Wavelength Range: |
450-1,000 nm
|
| Input Port diameter: |
4 mm
|
| Accuracy: |
± 0.5 nm (CW) ± 1.0 nm (pulsed)
|
| Resolution: |
0.1 nm (0.01 nm with HIRES option)
|
| Sensitivity: |
20 µW (CW @ 633 nm)10 µJ (pulsed)
|
| Dynamic Range: |
1000:1 with variable attenuator provided
|
| Damage Threshold (Attenuator.): |
75 W/cm2, 10 mJ/cm2 |
| CW Operation: |
CW or lasers modulated @ >5 kHz
|
| Warm-up Time: |
15 minutes
|
| Pulsed Operation: |
Single shot to <10 kHz Fully automatic triggering (pulse widths >100 f sec.) |
| Interface: |
RS-232 Serial Port (wavelength out)
|
| Power Supply: |
120 V/ 60 Hz, 240 V /50 Hz, rear panel selectable, fused 200/100 mA
|
| Weight: |
Sensor head, 0.56 kg, Control unit, 1.41 kg
|
| Mounting Post: |
0.250" 1/4-20 (6 mm thread also included)
|
| Operating Temperature: |
15-30 °C
|
| Dimensions: |
Control console, 155 x 70 x 205 mm, Sensor Head, 65 mm dia x 111 mm, with 32 x 82 mm attenuator block (pulsed operation only)
|
| Analog Output: |
0-1 V, calibrated proportional to wavelength (ANALOG option only)
|
|
|